Recall (RNA pathogen targets excluding rhinovirus and enterovirus; target level; reference: routine multiplex PCR) of Alternative post-hoc detection criterion (Sanderson et al. 2026) on Validation set: 344 prospectively collected nasopharyngeal swabs (Sanderson et al. 2026)

70% Recall (RNA pathogen targets excluding rhinovirus and enterovirus; target level; reference: routine multiplex PCR)

Methods

Tested configuration
Alternative post-hoc detection criterion (Sanderson et al. 2026)
Protocol
Target-level detection of respiratory pathogens against routine multiplex PCR, validation set (Sanderson et al. 2026)
Dataset
Validation set: 344 prospectively collected nasopharyngeal swabs (Sanderson et al. 2026)
Procedure
Target-level detection of respiratory pathogens against routine multiplex PCR, validation set (Sanderson et al. 2026)
Evaluation
Alternative post-hoc detection criterion on the validation set (Sanderson et al. 2026)
Coverage
Not reported scored / Not reported eligible
Uncertainty
95% CI 63 to 76
Evidence
Author-reported evaluation · source checkedClinical validation of a novel metagenomic nanopore sequencing method for detecting viral respiratory pathogens: diagnostic accuracy study · Results 'Performance by pathogen' paragraph 2 (after Figure 6 and its legend): "increased overall sensitivity using the alternative criteria further to 70% (145/207) (95% CI 63%-76%)"

A source-checked result verifies the numerical transcription, not every model or protocol detail. Evaluation metadata: source checked. Source checked does not mean independently reproduced.

Reproduction

Split
Validation set
Adaptation
Reads mapping to the target genome as a percentage of reads on the run mapping to that genome > 0.037%, AND >= 2 target reads, replacing the main and ratio criteria; chosen after the validation results
Scoring implementation
Target-level sensitivity and specificity against routine multiplex PCR

No execution recipe has been verified for this exact configuration and evaluation. A benchmark's general instructions may use different inputs, splits or model settings.

Reproducing this published result requires matching its model configuration, data, split and scorer. Source checking or a successful smoke test does not establish score reproduction.

Evidence

Source checking verifies the cited claim or transcription. It does not establish independent reproduction.

Evidence table

Inspect claims, sources and review details

Trace each statement to its source and review. A context-only reference supports the record generally; it does not verify an individual field. Source checking does not reproduce an experiment.

One row per statement and cited source. Multiple citations are not independent evaluations. Shared locators are labelled explicitly.

1 evidence row matching the loaded filters

Claims, original sources and review scope · Release 2026-10-22-1b7cf027e4f6
Property and statementOriginal source and locationReview and provenance
Reported result
70%
Individual claims
Clinical validation of a novel metagenomic nanopore sequencing method for detecting viral respiratory pathogens: diagnostic accuracy study

Original source ↗

Results 'Performance by pathogen' paragraph 2 (after Figure 6 and its legend): "increased overall sensitivity using the alternative criteria further to 70% (145/207) (95% CI 63%-76%)"

Version: medRxiv 10.64898/2026.02.06.26345651 v2, posted 2026-02-26; JATS XML
Retrieved: 2026-10-11T21:38:13Z

source checked

["source-hash-verification","independent-cell-check"] · 2026-10-11

author reported

Audit details

Re-downloaded the pinned artifact and matched its SHA-256; compared every value with a separate stdlib parser written for this review (the extractor's script was not run) and checked a stratified sample by hand against the rendered table.

Field: attributes.printed_value

Source artifact SHA-256: c0bc5c035742f53fe291e08b9d956102e102fd743a55de503f00bcaead36a7e2

Hash scope: Hash scope not separately documented; inspect source record

Inspected artifact

Extraction artifact SHA-256: c0bc5c035742f53fe291e08b9d956102e102fd743a55de503f00bcaead36a7e2

Extraction artifact

Sources and history

Release 2026-10-22-1b7cf027e4f6 · Record review: source checked

1 source record and release historyDownload this release (gzip)
Technical metadata and extraction receipts

Stable ID: rna-pathogen-20261011-sanderson2026-result-alternative-sensitivity-rna-excluding-rhino-entero

metric
recall
metric direction
higher
unit
percent
printed value
70%
numeric value
70
metric qualifier
RNA pathogen targets excluding rhinovirus and enterovirus; target level; reference: routine multiplex PCR
source locator
Results 'Performance by pathogen' paragraph 2 (after Figure 6 and its legend): "increased overall sensitivity using the alternative criteria further to 70% (145/207) (95% CI 63%-76%)"
numerator
145
denominator
207
uncertainty
type: confidence_interval; lower: 63; upper: 76; level: 0.95; printed: 95% CI 63%-76%
review
method: source-hash-verification; independent-cell-check; reviewer: claude; reviewer note: Separate Claude review agent (reviewer-path), independent of the extractor; no human review claimed; date: 2026-10-11; artifact sha256: c0bc5c035742f53fe291e08b9d956102e102fd743a55de503f00bcaead36a7e2; retrieval url: https://www.medrxiv.org/content/early/2026/02/26/2026.02.06.26345651.source.xml; method note: Transcribed from the Results text of the pinned JATS XML; the extraction script asserts that the quoted string occurs in the located paragraph.; note: Re-downloaded the pinned artifact and matched its SHA-256; compared every value with a separate stdlib parser written for this review (the extractor's script was not run) and checked a stratified sample by hand against the rendered table.